Rangger C, Daniel D M, Stone M L
Universitätsklinik für Unfallchirurgie Innsbruck.
Unfallchirurg. 1994 Sep;97(9):462-6.
The KT-1000 arthrometer was used to measure anterior tibial displacement in patients with acute unilateral anterior cruciate ligament (ACL) disruptions. Measurement revealed 11.5 mm absolute displacement on the injured side and 3.6 mm side-to-side difference with the 89 N test, 15.2 mm absolute displacement and 6.1 mm side-to-side difference with the manual maximum displacement test, and 8.5 mm displacement and 2.0 mm side-to-side difference with the active quadriceps test. Chronic anterior instabilities measured 13.0 mm absolute displacement on the involved side and 5.4 mm side-to-side difference with the 89N test, 17.7 mm displacement and 8.6 mm side-to-side difference with the manual maximum displacement test, 10.8 mm displacement and 4.4 mm side-to-side difference with the active quadriceps test. Testing technique, measurements in control subjects without previous knee injuries, and displacement measurements of other authors and other instrumented testing devices are described.
使用KT-1000关节测量仪测量急性单侧前交叉韧带(ACL)损伤患者的胫骨前移情况。测量结果显示,在89N试验中,患侧绝对位移为11.5mm,两侧差值为3.6mm;在手动最大位移试验中,绝对位移为15.2mm,两侧差值为6.1mm;在主动股四头肌试验中,位移为8.5mm,两侧差值为2.0mm。慢性前向不稳定患者在89N试验中,患侧绝对位移为13.0mm,两侧差值为5.4mm;在手动最大位移试验中,位移为17.7mm,两侧差值为8.6mm;在主动股四头肌试验中,位移为10.8mm,两侧差值为4.4mm。文中描述了测试技术、未受过膝关节损伤的对照受试者的测量结果,以及其他作者和其他仪器化测试设备的位移测量情况。