Effects of filtration, collimation, and target-receptor distance on artificial approximal enamel lesion detection with the use of RadioVisioGraphy.

作者信息

Parks E T, Miles D A, Van Dis M L, Williamson G F, Razmus T F, Bricker S L

机构信息

Department of Diagnostic Sciences, Indiana University School of Dentistry, Indianapolis.

出版信息

Oral Surg Oral Med Oral Pathol. 1994 Apr;77(4):419-26. doi: 10.1016/0030-4220(94)90207-0.

Abstract

RadioVisioGraphy is an imaging modality that uses a charge-coupled device electronic sensor. Dental charge-coupled device imaging can be used to detect dental caries; however, it was not known what effect beam collimation, added filtration, or variable target-receptor distance has on this task. The purpose of this investigation was to study these effects were imaged by conventional radiography and RadioVisioGraphy varying exposure time, target-receptor distance, collimation, and filtration. Printed RadioVisioGraphy images without enhancement and E-speed radiographs were evaluated by five viewers. The number of true-positives and false-positives were compared as were sensitivity, specificity, positive and negative predictive values, accuracy, and entrance level radiation exposure. Added filtration reduced the false-positives regardless of collimation or target-receptor distance. True-positives and false-positives were greater at a target-receptor distance of 8 inch at exposures < 0.10 seconds (F < 0.0151). E-speed radiographs had a greater specificity and positive predictive value than RadioVisioGraphy images without image enhancement; however, no significant differences were found between RadioVisioGraphy images and E-speed radiographs in true-positive or false-positive identifications without image enhancement.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验