Schwartz J G, Hurd I L, Carnahan J J
Department of Pathology, University of Texas Health Science Center at San Antonio 78284-7750.
Am J Emerg Med. 1994 Sep;12(5):513-6. doi: 10.1016/0735-6757(94)90266-6.
Both the Biosite Triage (Biosite Diagnostics, San Diego, CA) method and the Du Pont aca (Du Pont Company, Wilmington, DE) method give qualitative tricyclic antidepressant (TCA) results to aid in the diagnosis of a TCA overdose. The Triage method uses urine samples and the aca uses serum samples. Although the cutoff values vary considerably between the two methods, the Triage results agreed well with the aca results. The Triage test has an advantage in instrument maintenance and time savings, allowing a reduction in turn-around time for our emergency department. Both urine and serum samples were obtained from 44 patients who were admitted to the emergency department with a diagnosis of "possible tricyclic overdose." Discrepancies between the two methods were resolved by thin layer chromatography (Toxi-Lab, Ansys, Inc, Irvine, CA). Both methods were in agreement with the exception of five patients' samples. In this study, the Triage method allowed for detection of TCA using urine that is simple for the user and yielded higher sensitivity and specificity results compared with the Du Pont aca method.
Biosite Triage(Biosite诊断公司,加利福尼亚州圣地亚哥)方法和杜邦aca(杜邦公司,特拉华州威尔明顿)方法均可给出三环类抗抑郁药(TCA)的定性结果,以辅助TCA过量的诊断。Triage方法使用尿液样本,而aca方法使用血清样本。尽管两种方法的临界值差异很大,但Triage结果与aca结果吻合良好。Triage检测在仪器维护和节省时间方面具有优势,可缩短我们急诊科的周转时间。从44名因“可能三环类药物过量”诊断而入住急诊科的患者中获取了尿液和血清样本。两种方法之间的差异通过薄层色谱法(Toxi-Lab,Ansys公司,加利福尼亚州欧文)解决。除了五名患者的样本外,两种方法结果一致。在本研究中,Triage方法可使用尿液检测TCA,对用户来说操作简单,与杜邦aca方法相比,其灵敏度和特异性结果更高。