Verdonschot N J, Huiskes R, Freeman M A
Biomechanics Section, University of Nijmegen, The Netherlands.
Proc Inst Mech Eng H. 1993;207(3):149-54. doi: 10.1243/PIME_PROC_1993_207_287_02.
To investigate the accuracy of finite element (FE) models for pre-clinical testing of unbounded hip prostheses, relative to aspects of load transfer and micromobility, two previously published laboratory experiments were simulated, using three-dimensional FE models. It was found for the load-transfer analyses that the experiment and the FE study revealed results that were very similar. The trends in the mobility experiments were also reproduced in the FE simulations, although quantitative differences were found. It is concluded that FE analysis can effectively be used for design evaluation of hip prostheses before prototypes are made.
为了研究无约束髋关节假体临床前测试有限元(FE)模型在载荷传递和微动方面的准确性,使用三维有限元模型模拟了两个先前发表的实验室实验。对于载荷传递分析发现,实验和有限元研究得出的结果非常相似。虽然发现了定量差异,但在有限元模拟中也再现了微动实验的趋势。得出的结论是,在制造原型之前,有限元分析可有效地用于髋关节假体的设计评估。