[无骨水泥髋关节假体:PCA柄假体,5至7年随访结果]

[Cement-free hip endoprosthesis: PCA shaft prosthesis, 5-7-year results].

作者信息

Hösli P

机构信息

Orthopädische Universitätsklinik Basel.

出版信息

Z Orthop Ihre Grenzgeb. 1993 Nov-Dec;131(6):518-23. doi: 10.1055/s-2008-1040064.

Abstract

73 patients (out of a total of 79) with altogether 87 PCA-stems were followed up 5 to 7 years after total hip arthroplasty. The cups combined in these operations were not part of this study. In 11% the PCA-stem had to be removed respectively changed because of loosening. 50% of all patients suffered from thigh pain during the first 6 months postoperatively. Generally this pain disappeared within the first year after operation. Nevertheless 40% reported periods of transient thigh pain after the first postoperative year. Only when this pain was increasing and persisting over months, a correlation to loosening of the stem was found. Subsidence of more than 4 mm after the first postoperative year proved to be the only confident radiological sign of loosening. A strong correlation however was also found between appearance of a continuous radiolucent line through the zones 3-5 (according to Gruen) and loosening. We conclude that PCA-stems to our experience are not satisfactory because of the relatively high rate of loosening as well as the frequent occurring of transient thigh pain, even if this has not to be regarded as a sign of loosening.

摘要

79例患者中的73例共87个PCA柄在全髋关节置换术后5至7年接受了随访。这些手术中使用的髋臼杯不属于本研究范围。11%的PCA柄因松动而不得不被取出或更换。所有患者中有50%在术后前6个月出现大腿疼痛。一般来说,这种疼痛在术后第一年内消失。然而,40%的患者在术后第一年之后报告有短暂性大腿疼痛。只有当这种疼痛持续数月且加重时,才发现与柄的松动有关。术后第一年后下沉超过4毫米被证明是唯一可靠的松动放射学征象。不过,在Gruen分区3至5区出现连续的透亮线与松动之间也发现有很强的相关性。我们得出结论,根据我们的经验,PCA柄并不令人满意,因为其松动率相对较高,且经常出现短暂性大腿疼痛,即使这不被视为松动的迹象。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验