Okoshi T, Soldani G, Goddard M, Galletti P M
Artificial Organ Laboratory, Brown University, Providence, R.I.
J Thorac Cardiovasc Surg. 1993 May;105(5):791-5.
Two types of spongy polyurethane-polydimethylsiloxane blend (Cardiothane 51, Kontron Instruments, Inc., Everett, Mass.) vascular grafts with an internal diameter of 1.5 mm were fabricated by a spray, phase-inversion technique. Low-porosity grafts with hydraulic permeability of 2.7 +/- 0.4 ml/min per square centimeter and medium-porosity grafts with hydraulic permeability of 39 +/- 8 ml/min per square centimeter displayed good handling properties and suturability. Twelve straight low-porosity grafts, 17 straight medium-porosity grafts (1.5 to 2.0 cm in length), and one loop medium-porosity graft (10 cm in length) were implanted by the same surgeon end to end in the infrarenal aorta of 30 male Sprague-Dawley rats. Three months after implantation, patency was 8% for low-porosity grafts (1/12) and 76% for straight medium-porosity grafts (13/17). The loop medium-porosity graft was also patent. The sole patent low-porosity graft showed neointimal hyperplasia and incomplete endothelialization. All but one of the patent straight medium-porosity grafts showed a glistening and transparent neointima with complete endothelialization and no anastomotic hyperplasia. The loop medium-porosity graft displayed endothelialization from each anastomosis and in many islands in the middle portion of the graft, totalling 47% of the luminal surface by morphometric analysis. Thick mural thrombus, anastomotic hyperplasia, or aneurysm formation were not observed in any patent medium-porosity graft. These data indicate that in the rat aortic replacement model it is possible to achieve patency and a high degree of endothelialization in very small-diameter prostheses of appropriate porosity.
采用喷雾相转化技术制备了两种内径为1.5毫米的海绵状聚氨酯-聚二甲基硅氧烷共混物(Cardiothane 51,康创仪器公司,马萨诸塞州埃弗雷特)血管移植物。水力渗透率为2.7±0.4毫升/分钟每平方厘米的低孔隙率移植物和水力渗透率为39±8毫升/分钟每平方厘米的中孔隙率移植物表现出良好的操作性能和可缝合性。由同一位外科医生将12根直的低孔隙率移植物、17根直的中孔隙率移植物(长度为1.5至2.0厘米)和1根环状中孔隙率移植物(长度为10厘米)端端植入30只雄性Sprague-Dawley大鼠的肾下腹主动脉。植入后三个月,低孔隙率移植物的通畅率为8%(1/12),直的中孔隙率移植物的通畅率为76%(13/17)。环状中孔隙率移植物也保持通畅。唯一通畅的低孔隙率移植物显示有新生内膜增生和不完全内皮化。除一根外,所有通畅的直的中孔隙率移植物均显示有光滑透明的新生内膜,内皮化完全且无吻合口增生。通过形态计量分析,环状中孔隙率移植物在每个吻合口以及移植物中部的许多区域均有内皮化,占管腔表面的47%。在任何通畅的中孔隙率移植物中均未观察到厚壁血栓、吻合口增生或动脉瘤形成。这些数据表明,在大鼠主动脉置换模型中,对于具有适当孔隙率的非常小直径的假体,有可能实现通畅和高度内皮化。