Intracardiac echocardiography: in vitro and in vivo validation for right ventricular volume and function.

作者信息

Vazquez de Prada J A, Chen M H, Guerrero J L, Padial L R, Jiang L, Schwammenthal E, Sagie A, Weyman A E, Levine R A, Chen C

机构信息

Cardiac Ultrasound Laboratory, Massachusetts General Hospital, Boston, USA.

出版信息

Am Heart J. 1996 Feb;131(2):320-8. doi: 10.1016/s0002-8703(96)90361-0.

Abstract

To determine the feasibility and accuracy of intracardiac ultrasonography (ICUS) for the measurement of right ventricular (RV) volumes and function, a 10 MHz ICUS catheter was used in an in vitro and in vivo model. In the in vitro study, 16 sheep hearts were imaged. Sequential cross-sectional images from RV apex to base were recorded during a calibrated pullback. Volumes were calculated by applying Simpson's algorithm. ICUS-obtained volumes correlated well with actual volumes (standard error of estimate [SEE] = 2.3 ml, r = 0.98). For the in vivo study, a beating-heart canine model was used (31 hemodynamic stages in six dogs). Actual volumes were measured by an intracavitary balloon connected to an external column. Sequential cross-sectional images were recorded during the ICUS catheter pullback from apex to base of the RV, and volumes calculated by Simpson's algorithm. Good correlations were observed between ICUS and actual values for diastolic (SEE = 4.1 ml, r = 0.97), systolic (SEE = 3.4 ml, r = 0.96), and ejection fraction (SEE = 3.1%, r = 0.87) values. This new technique can accurately quantitate RV volumes, can function both in vitro and in vivo, and has the potential for increasing applications to questions of clinical and research interest.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验