在Cobas Mira S分析仪上对血红蛋白A1c免疫比浊法的评估。

Evaluation of an immunoturbidimetric assay for haemoglobin A1c on a Cobas Mira S analyser.

作者信息

Weets I, Gorus F K, Gerlo E

机构信息

Laboratorium Klinische Chemie, Akademisch Ziekenhuis, Vrije Universiteit Brussel, Belgium.

出版信息

Eur J Clin Chem Clin Biochem. 1996 May;34(5):449-53. doi: 10.1515/cclm.1996.34.5.449.

Abstract

We evaluated a homogenous immunoturbidimetric assay for haemoglobin A1c (Tina-quant Haemog bin A1c, Boehringer Mannheim, GmbH, Mannheim, Germany) adapted to a Cobas Mira S analyser (F. Hoffmann-La Roche & Co., Basel, Switzerland) and not requiring sample pretreatment. Between-day CV's determined over a 6 week period were 5.9% and 5.3% for mean haemoglobin A1c values of 5.4% and 17.0% of total haemoglobin respectively. The imprecision was higher than with an in-house ion-exchange high performance liquid chromatographic method. Bilirubin, triacylglycerols and the labile fraction of haemoglobin A1c did not interfere with the assay of haemoglobin A1c. Fetal haemoglobin is not recognized by the antibodies used. Results correlated well with those obtained by high performance liquid chromatography. In conclusion, the Tina-quant assay is not prone to common interferences and allows the rapid and automated determination of haemoglobin A1c on open photometric analysers.

摘要

我们评估了一种用于糖化血红蛋白的均相免疫比浊法(Tina-quant糖化血红蛋白检测试剂盒,德国曼海姆市宝灵曼有限公司),该方法适用于Cobas Mira S分析仪(瑞士巴塞尔市霍夫曼-罗氏有限公司),且无需对样品进行预处理。在为期6周的时间里,分别针对糖化血红蛋白平均含量为总血红蛋白5.4%和17.0%的情况测定日间变异系数,结果分别为5.9%和5.3%。其不精密度高于实验室内部的离子交换高效液相色谱法。胆红素、三酰甘油以及糖化血红蛋白的不稳定部分均不干扰糖化血红蛋白的检测。所用抗体无法识别胎儿血红蛋白。检测结果与高效液相色谱法所得结果相关性良好。总之,Tina-quant检测法不易受到常见干扰因素的影响,可在开放式光度分析仪上快速、自动地测定糖化血红蛋白。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验