Zhang S S, Liu H X, Chen Y, Yuan Z B
Center of Instrumental Analysis, Zhengzhou University, P.R. China.
Biomed Chromatogr. 1996 Sep-Oct;10(5):256-7. doi: 10.1002/(SICI)1099-0801(199609)10:5<256::AID-BMC600>3.0.CO;2-C.
High-performance capillary electrophoresis (HPCE) and high-performance liquid chromatography (HPLC) were developed and applied to the determination of acyclovir (ACV) and guanine (G) in pharmaceuticals. The comparison study showed that two methods gave comparable results in linear range, recovery and reproducibility. HPCE was used for the determination of ACV and G in urine; the recovery was better than 81.3% and the RSD was less than 4.4%.
高效毛细管电泳(HPCE)和高效液相色谱(HPLC)被开发并应用于药物中阿昔洛韦(ACV)和鸟嘌呤(G)的测定。对比研究表明,两种方法在线性范围、回收率和重现性方面给出了可比的结果。HPCE用于尿液中ACV和G的测定;回收率优于81.3%,相对标准偏差小于4.4%。