The effect of a soft collar, used as normally recommended or reversed, on three planes of cervical range of motion.

作者信息

Carter V M, Fasen J A, Roman J M, Hayes K W, Petersen C M

机构信息

TheraTx, Denver, CO, USA.

出版信息

J Orthop Sports Phys Ther. 1996 Mar;23(3):209-15. doi: 10.2519/jospt.1996.23.3.209.

Abstract

Clinicians recommend soft cervical collars to immobilize the cervical spine following trauma. They apply them either as intended by the manufacturer or reversed (collar rotated 180 degrees), purportedly to achieve limitation in a specific direction. This study investigated the effectiveness of soft cervical collars in limiting cervical range of motion when worn as recommended or reversed. All planes of cervical range of motion of 50 volunteer subjects without current or past cervical dysfunction were measured under three conditions (no collar, recommended use, and reversed) using the Orthopedic Systems Inc. Computerized Anatometry-6000 Spine Motion Analyzer. The instrument has been previously shown to produce measures with high reliability and to correlate strongly with known angular measures. Within subjects analyses of variance indicated significant differences in all six ranges of motions among the three conditions. Post hoc paired t tests showed that wearing a collar either as recommended or reversed decreased motion compared with not wearing a collar, and that the position of the collar affected range of motion in three of the six motions. Differences in range can be attributed to location of the collar closure and initial head posture. Soft cervical collars can physically limit motion when worn either way.

摘要

临床医生建议在创伤后使用软质颈托固定颈椎。他们要么按照制造商的意图使用颈托,要么将其翻转(颈托旋转180度),据称这样能在特定方向上限制活动。本研究调查了软质颈托按推荐方式佩戴或翻转佩戴时限制颈椎活动范围的有效性。使用Orthopedic Systems Inc.的计算机化解剖学6000脊柱运动分析仪,在三种情况下(不佩戴颈托、按推荐方式佩戴、翻转佩戴)测量了50名无当前或既往颈椎功能障碍的志愿者受试者颈椎活动范围的所有平面。该仪器先前已被证明能产生可靠性高的测量结果,且与已知角度测量结果高度相关。受试者内方差分析表明,三种情况下所有六个活动范围均存在显著差异。事后配对t检验表明,与不佩戴颈托相比,按推荐方式佩戴或翻转佩戴颈托都会减少活动,且颈托的位置在六个活动中的三个中影响活动范围。活动范围的差异可归因于颈托闭合处的位置和初始头部姿势。软质颈托无论以哪种方式佩戴都能在物理上限制活动。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验