Abstract

An anthropomorphic phantom was used to calibrate a supine geometry sodium iodide total body potassium monitor. Correction factors accommodating variability in subject size were empirically determined. Measurements on 12 males of weight 45-96 kg, height 161-184 cm and 18 females of weight 48-89 kg, height 153-175 cm, showed that the 40K calibration factor (F, counts s-1 (g of potassium)-1) was significantly correlated (r = 0.88, p < 0.0001) to subject (weight/height)0.5, indicating comparable accuracy to 42K-based calibration procedures. Fat-free mass determined from the potassium measurements of 16 subjects correlated significantly with fat-free mass estimated from skinfold thickness (r = 0.98, p < 0.0001), dual-energy x-ray absorptiometry (r = 0.99, p < 0.0001) and bioimpedance analysis (r = 0.98, p < 0.0001). These data, together with the precision (coefficient of variation, CV = 1.5%) and accuracy (CV = 4.5%) of the system, indicate that this calibration procedure represents a relatively low-cost, non-invasive alternative to 42K-based methods of calibrating total body potassium monitors.

摘要

使用一个人体模型来校准仰卧位几何碘化钠全身钾监测仪。通过实验确定了适应受试者体型变化的校正因子。对12名体重45 - 96千克、身高161 - 184厘米的男性和18名体重48 - 89千克、身高153 - 175厘米的女性进行测量,结果显示40K校准因子(F,计数·秒⁻¹·(每克钾)⁻¹)与受试者(体重/身高)⁰·⁵显著相关(r = 0.88,p < 0.0001),表明其准确性与基于42K的校准程序相当。根据16名受试者的钾测量值确定的去脂体重与通过皮褶厚度估计的去脂体重(r = 0.98,p < 0.0001)、双能X射线吸收法(r = 0.99,p < 0.0001)和生物电阻抗分析(r = 0.98,p < 0.0001)显著相关。这些数据,连同该系统的精密度(变异系数,CV = 1.5%)和准确度(CV = 4.5%),表明这种校准程序是一种相对低成本、非侵入性的替代基于42K的全身钾监测仪校准方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验