McCall R C
McCall Associates, Woodside, California 94062, USA.
Med Phys. 1997 Jan;24(1):135-6. doi: 10.1118/1.597914.
The problem of calculating the neutron capture gamma-ray dose rate due to thermal neutron capture in a boron or cadmium rectangular shield is considered. An example is given for shielding for a door at the exit of medical accelerator room maze in order to determine the optimum location of lead relative to the borated polyethylene.