Abstract

We describe positive interference with the ion-selective electrode determination of lithium (Lytening 2Z analyzer; Dade) when blood is collected in a 10-mL plain red-top plastic Vacutainer Plus Tube (Becton Dickinson) containing a silica clot activator and silicone surfactant (prod. no. 36-7820). We evaluated both the original tube (blue-labeled) and a new tube formulated to contain less silicone surfactant (striped-labeled). We determined that the interference is from either the silica clot activator or the silicone surfactant used to fix the silica to the tube and is inversely related to the volume of blood in the tube. Long-term intermittent exposure of the Li ion-selective electrode to the silica clot activator or surfactant results in decreased Li values--in terms of both the positive interference by the silica clot activator or surfactant and the actual Li determinations. Moreover, this long-term interference with the Li ion-selective electrode for patient's specimens is undetected by the Dade control material (QCLytes).

摘要

我们描述了在使用离子选择电极测定锂(Lytening 2Z分析仪;达德公司)时,当血液收集在含有硅胶促凝剂和硅氧烷表面活性剂的10毫升普通红色顶盖塑料真空采血管(Becton Dickinson公司,产品编号36 - 7820)中时出现的正干扰。我们评估了原始管(蓝色标签)和一种新配方的管,该新管含有较少的硅氧烷表面活性剂(条纹标签)。我们确定干扰来自硅胶促凝剂或用于将硅胶固定在管上的硅氧烷表面活性剂,并且与管中血液的体积成反比。锂离子选择电极长期间歇性暴露于硅胶促凝剂或表面活性剂会导致锂值降低——这既涉及硅胶促凝剂或表面活性剂的正干扰,也涉及实际的锂测定。此外,达德公司的质控材料(QCLytes)未检测到对患者样本的锂离子选择电极的这种长期干扰。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验