Suppr超能文献

Malfunctioning Starr-Edwards mitral valve 21 years after installation.

作者信息

Sakata K, Ishikawa S, Ohtaki A, Otani Y, Suzuki M, Kawashima O, Morishita Y

机构信息

Second Department of Surgery, Gunma University School of Medicine, Maebashi, Japan.

出版信息

J Cardiovasc Surg (Torino). 1997 Feb;38(1):81-2.

PMID:9128128
Abstract

Two cases of malfunctioning Starr-Edwards cloth-covered mitral valve prostheses requiring reoperation are presented. Both cases underwent successful surgical repair 21 years after the valve replacement. The causes were a disturbance of the poppet during the opening movement due to excessive tissue ingrowth and a paravalvular leak associated with a tear of the valve seat. Replacement of the Starr-Edwards valve prosthesis more than 20 years after the initial installation has not been reported.