Two-dimensional Fourier filtration of acoustic quantification echocardiographic images: improved reproducibility and accuracy of automated measurements of left ventricular performance.

作者信息

Chandra S, Garcia M J, Morehead A, Thomas J D

机构信息

Department of Cardiology, Cleveland Clinic Foundation, Veterans Affairs Medical Center, OH 044195, USA.

出版信息

J Am Soc Echocardiogr. 1997 May;10(4):310-9. doi: 10.1016/s0894-7317(97)70067-1.

Abstract

To determine the accuracy of Fourier filtration in removing the high-frequency component of noise from acoustic quantification (AQ) echocardiographic images, we processed 800 parasternal short-axis images obtained from 10 study subjects. M-mode tracings were also obtained and used as gold standard for correlating the results from raw AQ and Fourier-filtered images. Left ventricular short-axis diameters measured from the raw AQ and Fourier-filtered data were compared with the M-mode diameters (r = 0.91, p < 0.001 for raw AQ; and r = 0.96, p < 0.001, for Fourier filtered images). Fractional shortening showed better correlation between Fourier-filtered images and M-mode (r = 0.79, p < 0.03) versus raw AQ and M-mode (r = 0.33, p = 0.46). Best-to-beat reproducibility was also found to be better for fractional area change (r = 0.82, p = 0.01 versus r = 0.66, p = 0.77), peak area filling rate (r = 0.87, p = 0.004 versus r = 0.62, p = 0.1), and peak are emptying rate (r = 0.99, p < 0.0001 versus r = 0.19, p < 0.7) for Fourier filtered versus raw AQ. Our results indicate that Fourier filtration of AQ data results in more accurate representation of the true endocardial borders.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验