Lentjes E G, Romijn F P, Moolenaar A J
Department of Clinical Chemistry, University Hospital Leiden, The Netherlands.
J Steroid Biochem Mol Biol. 1997 Feb;60(3-4):255-60. doi: 10.1016/s0960-0760(96)00186-0.
We have determined the free cortisol concentration in serum using either the Amicon MPS-1 ultrafiltration-centrifugation method (I) or equilibrium dialysis (II). If procedure I was used we found that [1,2,6,7-3H]-, and [4-14C]cortisol had a lower affinity than unlabelled cortisol for corticosteroid binding globulin (CBG). The binding affinity (Ka) to three separate CBG-containing samples was 8-18 times lower for [1,2,6,7-3H]cortisol and 30-90 times lower for [4-14C]cortisol, when compared with that of unlabelled cortisol. This difference in affinity to CBG was not observed if method II was used for the free cortisol determinations. The observed isotope effect in method I is not caused by unspecific binding to material such as the Amicon MPS-1 chamber or to impurities in the tracer. We suggest that the centrifugation step during ultrafiltration changed the conformation of CBG, thereby reducing its affinity for labelled cortisol. It is concluded that incorrect results will be obtained if radiolabelled is cortisol used for determining the free cortisol content of plasma with the Amicon MPS-1 device.
我们使用Amicon MPS - 1超滤离心法(I)或平衡透析法(II)测定了血清中的游离皮质醇浓度。如果使用方法I,我们发现[1,2,6,7 - 3H]-皮质醇和[4 - 14C]皮质醇与皮质类固醇结合球蛋白(CBG)的亲和力低于未标记的皮质醇。与未标记的皮质醇相比,[1,2,6,7 - 3H]皮质醇与三个不同含CBG样品的结合亲和力(Ka)低8 - 18倍,[4 - 14C]皮质醇低30 - 90倍。如果使用方法II测定游离皮质醇,则未观察到对CBG亲和力的这种差异。方法I中观察到的同位素效应不是由与诸如Amicon MPS - 1腔室等材料或示踪剂中的杂质的非特异性结合引起的。我们认为超滤过程中的离心步骤改变了CBG的构象,从而降低了其对标记皮质醇的亲和力。结论是,如果使用放射性标记的皮质醇通过Amicon MPS - 1装置测定血浆中的游离皮质醇含量,将获得错误的结果。