Feasibility of ultrasound hyperthermia with waveguide interstitial applicator.

作者信息

Jarosz B J

机构信息

Ottawa-Carleton Institute for Physics, Department of Physics, Carleton University, Canada.

出版信息

IEEE Trans Biomed Eng. 1996 Nov;43(11):1106-15. doi: 10.1109/10.541252.

Abstract

Interstitial and intracavitary ultrasonic hyperthermia applicators facilitate well-controlled power deposition in tissues. In this paper, analysis of temperature elevation and experimental results in tissue phantom, animal tissue in vivo and animal tissue in vitro are presented for a waveguide applicator intended for treatment of brain tumors. It consisted of a G18 hypodermic needle attached via a conical velocity transformer to a 12.7-mm-diameter piezoelectric disk operated at 1.0 MHz. The axial acoustic pressure distribution had a standing-wave pattern with the four cycles/cm spatial periodicity. This periodicity was absent in the temperature distribution in tissue phantoms. The simulations based on a solution to the effective heat conductivity equation indicated that the hyperthermic range can be reached within a 4- and a 10-mm radius around the applicator for a 21- and a 60-mm sample diameters, respectively, with reasonable input power. The first number corresponded closely to the 5-mm radius observed in porcine brain in vivo and the second one came close to the 9-mm radius in porcine brain in vitro. The presented results demonstrate the potential of the ultrasound waveguide interstitial applicator for hyperthermia of small volume tumors.