Blick D W, Adair E R, Hurt W D, Sherry C J, Walters T J, Merritt J H
Systems Research Laboratories, Inc., Brooks AFB, Texas 78235, USA.
Bioelectromagnetics. 1997;18(6):403-9.
We measured thresholds for microwave-evoked skin sensations of warmth at frequencies of 2.45, 7.5, 10, 35, and 94 GHz. In the same subjects, thresholds of warmth evoked by infrared radiation (IR) were also measured for comparison. Detection thresholds were measured on the skin in the middle of the back in 15 adult male human subjects at all microwave (MW) frequencies and with IR. Long duration (10-s), large area (327-cm2) stimuli were used to minimize any differential effects of temporal or spatial summation. Sensitivity increased monotonically with frequency throughout the range of microwave frequencies tested. The threshold at 94 GHz (4.5 +/- 0.6 mW/cm2) was more than an order of magnitude less than at 2.45 GHz (63.1 +/- 6.7 mW/cm2), and it was comparable to the threshold for IR (5.34 +/- 1.07 mW/cm2).
我们测量了2.45、7.5、10、35和94吉赫兹频率下微波诱发皮肤温热感觉的阈值。对同一批受试者,还测量了红外辐射(IR)诱发的温热阈值以作比较。在15名成年男性受试者的背部中央皮肤上测量了所有微波(MW)频率以及红外辐射下的检测阈值。使用持续时间长(10秒)、面积大(327平方厘米)的刺激来最小化时间或空间总和的任何差异效应。在所测试的整个微波频率范围内,敏感度随频率单调增加。94吉赫兹时的阈值(4.5±0.6毫瓦/平方厘米)比2.45吉赫兹时的阈值(63.1±6.7毫瓦/平方厘米)低一个多数量级,且与红外辐射的阈值(5.34±1.07毫瓦/平方厘米)相当。