Komatani A, Hasegawa N, Yamaguchi K, Manaka Y, Takahashi K
Department of Radiology, Yamagata University School of Medicine.
Kaku Igaku. 1997 Jun;34(6):385-8.
We previously reported about the room air contamination during ventilation study using the TECHNEGAS system. Main cause of the contamination with radioactive gas were leakage through the filter and expiration of the patient after the inhalation. Then, a new type filter was developed to reduce the radioactive leakage. The efficacy and usefulness of the new type filter was examined. The radioactive leakage through the new filter was markedly reduced to 1/6 compared with the old type. But the air resistance of passage was increased over two times of the old type. This resistance is not able to disregard on respiratory physiology. Improvement to reduce the resistance is desirable.
我们之前报道过使用锝气体(TECHNEGAS)系统进行通气研究期间室内空气污染的情况。放射性气体污染的主要原因是过滤器泄漏以及患者吸入后呼出。随后,研发了一种新型过滤器以减少放射性泄漏。对新型过滤器的功效和实用性进行了检测。与旧型过滤器相比,新型过滤器的放射性泄漏显著降低至原来的1/6。但是气体通过的阻力增加到旧型过滤器的两倍多。这种阻力在呼吸生理学上不容忽视。减少阻力的改进是可取的。