De Wachter D S, Verdonck P R, De Vos J Y, Hombrouckx R O
Hydraulics Laboratory, Institute Biomedical Technology, University of Gent, Belgium.
Int J Artif Organs. 1997 Jul;20(7):366-70.
To investigate the haemolysis in haemodialysis cannulae, an in-vitro set up is built, using a unipuncture dialysis system. This system is connected to a bag with fresh calf's blood, by the cannula under test, mounted in a large bloodline (5 mm diameter). The blood characteristics are kept constant by means of a bicarbonate dialysate in the dialyser. During a 6 h period, haematological parameters are regularly samples. Flow through the cannulae is recorded, which is about 500 mL/min. Four different cannulae are tested and compared to the results obtained without any cannula in the circuit. In all cases a linear increase in plasma free haemoglobin levels is found after 6 h. The cannulae can be ranked from 8F catheter over 13G, 14G to 16G cannula, the latter producing the highest degree of haemolysis. When using plastic cannulae at high blood flows, their haemolytic effect may not be neglected.
为了研究血液透析套管中的溶血情况,利用单针透析系统建立了一个体外装置。该系统通过受试套管连接到装有新鲜小牛血液的袋子上,受试套管安装在一个大的血路(直径5毫米)中。通过透析器中的碳酸氢盐透析液使血液特性保持恒定。在6小时期间,定期采集血液学参数样本。记录通过套管的血流量,约为500毫升/分钟。测试了四种不同的套管,并将结果与回路中不使用任何套管时获得的结果进行比较。在所有情况下,6小时后血浆游离血红蛋白水平均呈线性增加。套管可按从8F导管到13G、14G再到16G套管的顺序排列,16G套管产生的溶血程度最高。在高血流量下使用塑料套管时,其溶血作用不可忽视。