Intratumoral vascularity of experimentally induced VX2 carcinoma: comparison of color Doppler sonography, power Doppler sonography, and microangiography.

作者信息

Park K S, Choi B I, Won H J, Seo J B, Kim S H, Kim T K, Han J K, Yeon K M

机构信息

Department of Diagnostic Radiology, Chung Buk National University College of Medicine, Korea.

出版信息

Invest Radiol. 1998 Jan;33(1):39-44. doi: 10.1097/00004424-199801000-00006.

Abstract

RATIONALE AND OBJECTIVES

Tumor vascularity is useful for characterizing tumors and determining tumor management. The recent development of power Doppler sonography has enhanced the sensitivity of color Doppler imaging in the detection of blood flow because of low power noise and less angle dependence. The purpose of this study was to compare the capability of color and power Doppler sonography with that of microangiography for showing tumor vascularity of VX2 carcinoma.

METHODS

Color and power Doppler sonography was performed on VX2 carcinomas in the rabbit thighs, and their findings were correlated with those of microangiography. For qualitative analysis, tumor vascularity was categorized into four items including distribution of tumor vessels, crowdedness of vessels, small vessels, (> or = 0.1 mm), and micro-vessels (< 0.1 mm). Tumor blood flow signals of color Doppler sonography and power Doppler sonography were graded as 3, 2, 1, and 0 and were compared with tumor vascularity on microangiography. For quantitative analysis, percentages of tumor vascular area per tumor area on each study were compared.

RESULTS

The mean scores of tumor vascularity on power Doppler sonography were 2.87, 2.73, 2.93, and 2.73 in tumor vascular distribution, crowdedness of vessels, small vessels, and micro-vessels, respectively. Those on color Doppler sonography were 2.4, 2.2, 2.8, and 1.67, respectively. Power Doppler sonography was statistically superior to color Doppler sonography in displaying tumor vascular distribution (P < 0.05) and micro-vessels (P < 0.01). The means and medians of percentages of tumor vascular area per tumor area were 22.7% and 23.5% on microangiography, 17.9% and 21.4% on color Doppler sonography, and 36.4% and 34.7 % on power Doppler sonography, respectively. Percentages of tumor vascular area per tumor area on both color Doppler sonography (r = 0.70) and power Doppler sonography (r = 0.84) were well correlated with those on microangiography.

CONCLUSIONS

Power Doppler sonography can demonstrate the tumor vascularity on microangiography relatively well, however, it tends to overestimate the blood flow signals. Color Doppler sonography may have some limitations in imaging tumor vascular distribution and micro-vessels and tends to underestimate tumor vascularity, but can display the blood flow direction. Therefore, power Doppler and color Doppler sonography could complement each other in demonstrating the tumor vascularity.

摘要

原理与目的

肿瘤血管生成有助于肿瘤特征的描述及治疗方案的确定。由于低功率噪声和较小的角度依赖性,能量多普勒超声的最新进展提高了彩色多普勒成像检测血流的敏感性。本研究的目的是比较彩色多普勒超声和能量多普勒超声与微血管造影显示VX2癌肿瘤血管生成的能力。

方法

对兔大腿的VX2癌进行彩色多普勒超声和能量多普勒超声检查,并将其结果与微血管造影结果进行对比。进行定性分析时,将肿瘤血管生成分为四项,包括肿瘤血管分布、血管密集程度、小血管(≥0.1mm)和微血管(<0.1mm)。将彩色多普勒超声和能量多普勒超声的肿瘤血流信号分为3、2、1和0级,并与微血管造影的肿瘤血管生成情况进行比较。进行定量分析时,比较各项研究中肿瘤血管面积占肿瘤总面积的百分比。

结果

能量多普勒超声在肿瘤血管分布、血管密集程度、小血管和微血管方面的肿瘤血管生成平均得分分别为2.87、2.73、2.93和2.73。彩色多普勒超声的得分分别为2.4、2.2、2.8和1.67。在显示肿瘤血管分布(P<0.05)和微血管(P<0.01)方面,能量多普勒超声在统计学上优于彩色多普勒超声。微血管造影中肿瘤血管面积占肿瘤总面积的百分比的均值和中位数分别为22.7%和23.5%,彩色多普勒超声为17.9%和21.4%,能量多普勒超声为36.4%和34.7%。彩色多普勒超声(r=0.70)和能量多普勒超声(r=0.84)的肿瘤血管面积占肿瘤总面积的百分比与微血管造影的结果均具有良好的相关性。

结论

能量多普勒超声能够较好地显示微血管造影上的肿瘤血管生成情况,然而,它往往会高估血流信号。彩色多普勒超声在显示肿瘤血管分布和微血管方面可能存在一些局限性,并且往往会低估肿瘤血管生成,但能够显示血流方向。因此,能量多普勒超声和彩色多普勒超声在显示肿瘤血管生成方面可以相互补充。