Fernández-Abedul M T, Costa-García A
Departamento de Química Física y Analítica, Universidad de Oviedo, Asturias, Spain.
J Pharm Biomed Anal. 1997 Sep;16(1):15-9. doi: 10.1016/s0731-7085(96)01962-0.
Flow injection analysis (FIA) with amperometric detection using a carbon paste electrode is applied to the determination of naltrexone. The sample solution was injected into the carrier stream of 0.1 M perchloric acid, being determined by oxidation at +1.0 V vs. Ag/AgCl/sat. KCl using a flow rate of 4 ml min-1. A relative standard deviation of 1.5% was calculated for a concentration level of 10(-5) M (n = 17) without carrying out a carbon paste electrode pretreatment. Calibration curves were found to be linear between 2 x 10(-8) and 10(-5) M (almost three orders of magnitude) and the method has a detection limit of 2 x 10(-8) M. A simple and reproducible procedure is proposed for the determination of naltrexone in pharmaceuticals. The results compared favourably with those obtained by an HPLC-UV method.
采用碳糊电极安培检测的流动注射分析(FIA)用于测定纳曲酮。将样品溶液注入0.1 M高氯酸载流中,以相对于Ag/AgCl/饱和KCl在+1.0 V下氧化进行测定,流速为4 ml min-1。在不进行碳糊电极预处理的情况下,对于10(-5) M的浓度水平计算出相对标准偏差为1.5%(n = 17)。发现校准曲线在2 x 10(-8)至10(-5) M之间呈线性(几乎三个数量级),该方法的检测限为2 x 10(-8) M。提出了一种用于测定药物中纳曲酮的简单且可重复的方法。结果与通过HPLC-UV法获得的结果相比具有优势。