Suppr超能文献

诊断X射线光谱:不同计算方法生成的光谱与实测光谱的比较。

Diagnostic x-ray spectra: a comparison of spectra generated by different computational methods with a measured spectrum.

作者信息

Bhat M, Pattison J, Bibbo G, Caon M

机构信息

School of Physics and Electronic Systems Engineering, University of South Australia, The Levels.

出版信息

Med Phys. 1998 Jan;25(1):114-20. doi: 10.1118/1.598170.

Abstract

A number of computer codes, developed using semi-empirical models, are available to compute x-ray spectra from a tungsten target for different tube parameters. In this study x-ray spectra measured with a high-purity germanium detector are compared with those computed using the empirical models and previously published measured data. The computer codes used to generate the spectra are based on models proposed by Birch et al. and Tucker et al. The measured x-ray spectra agreed well with the computed x-ray spectra using the model of Tucker et al. whereas the model of Birch et al. produced a "harder" x-ray spectrum compared to the measured spectra. Our measured x-ray spectra compared well with the previously published measured spectral data of Fewell et al.

摘要

有许多使用半经验模型开发的计算机代码,可用于计算不同管参数下钨靶的X射线光谱。在本研究中,将用高纯锗探测器测量的X射线光谱与使用经验模型计算的光谱以及先前发表的测量数据进行了比较。用于生成光谱的计算机代码基于Birch等人和Tucker等人提出的模型。测量的X射线光谱与使用Tucker等人的模型计算的X射线光谱吻合良好,而Birch等人的模型产生的X射线光谱比测量光谱“更硬”。我们测量的X射线光谱与Fewell等人先前发表的测量光谱数据比较吻合。