Wang K, Garcia F, Leonelli F M
Division of Cardiovascular Medicine, University of Kentucky Medical Center, Lexington, USA.
G Ital Cardiol. 1998 Feb;28(2):97-101.
This study evaluates the immediate effects of the endocardial electrical shocks delivered by a transvenous defibrillation system on left ventricular (LV) function in a pig model. A triple-lead system consisting of two endocardial electrodes, in the right ventricular apex and the junction of superior cava-right atrium, and a custom-made defibrillation can implanted subcutaneously in the thorax was set up in 10 close-chest pigs. Transesophageal echocardiography with two dimensional image, m-Mode, and pulse Doppler was performed at baseline and after several episodes of fibrillation/defibrillation (F/DF). Each animal underwent an average of 8 (range 6 to 11) episodes of ventricle F/DF for a total of 210 (range 165 to 290) joules of biphasic-waveform defibrillation shocks. Heart rate, blood pressure, LV end-systolic area, end-diastolic area and fractional area contraction, isovolumic relaxation time, and both ratios of velocities and time-velocity integrals in transmitral Doppler flow E and A waves were unchanged after the shocks. This animal study suggests that multiple countershocks up to 210 joules delivered by a transvenous defibrillation system do not cause LV global systolic and/or diastolic dysfunction.
本研究评估经静脉除颤系统施加的心内膜电冲击对猪模型左心室(LV)功能的即时影响。在10只开胸猪中建立了一个三导联系统,该系统由两个心内膜电极组成,分别置于右心室心尖和上腔静脉-右心房交界处,以及一个定制的除颤罐,皮下植入胸部。在基线以及几次颤动/除颤(F/DF)发作后,采用二维图像、M型和脉冲多普勒经食管超声心动图检查。每只动物平均经历8次(范围6至11次)心室F/DF发作,总共接受210焦耳(范围165至290焦耳)的双相波形除颤电击。电击后,心率、血压、左心室收缩末期面积、舒张末期面积和面积缩短分数、等容舒张时间以及二尖瓣多普勒血流E波和A波的速度比值和时间-速度积分均未改变。这项动物研究表明,经静脉除颤系统施加的高达210焦耳的多次反搏电击不会导致左心室整体收缩和/或舒张功能障碍。