Shibata Y, Doi O, Goto T, Hase T, Kadota K, Fujii M, Zenke M, Fujii S, Ashida N, Sugioka J, Yamamoto H, Nishizaki M, Kameko M, Mitsudo K
Department of Cardiology, Kurashiki Central Hospital, Okayama, Japan.
Cathet Cardiovasc Diagn. 1998 Mar;43(3):344-51. doi: 10.1002/(sici)1097-0304(199803)43:3<344::aid-ccd24>3.0.co;2-k.
A new guiding catheter for PTCA is described. In our department, 302 patients (405 lesions) underwent transradial coronary angioplasty using the 6 Fr Kimny guiding catheter since January 1996. The total engagement rate using the Kimny guiding catheter was 91.3% (370/405). The engagement rate after the modified Kimny guiding catheter was introduced in May 1996 increased to 96.0% (243/253). The stent delivery success rate was 98.4%. We had two dislodged stents. PTCA for both left and right coronary arteries in a single procedure with the Kimny guiding catheter was performed via the radial artery in 27 patients. In 24 of these patients (89%) we engaged both coronaries successfully. In the remaining 3 patients we switched to another catheter. Except for 4 patients with non-Q-wave myocardial infarction, no major cardiac complications were encountered. No major entry site-related complications were seen, and no patient required vascular surgery or blood transfusions. In one patient the Kimny guiding catheter tip caused a minor dissection of the LMT, but no ischemic event occurred as a result. In conclusion, the Kimny device is a useful PTCA guiding catheter for routine angioplasty and stenting.