Thanopoulos B D, Laskari C V, Tsaousis G S, Zarayelyan A, Vekiou A, Papadopoulos G S
Department of Pediatric Cardiology, Aghia Sophia Children's Hospital, Athens, Greece.
J Am Coll Cardiol. 1998 Apr;31(5):1110-6. doi: 10.1016/s0735-1097(98)00039-4.
This study reports our clinical experience with transcatheter closure of secundum atrial septal defects (ASDs) in children, using the Amplatzer, a new occlusion device.
None of the devices previously used for transcatheter closure of interatrial communications has gained wide acceptance.
We examined the efficacy and safety of the Amplatzer, a new self-centering septal occluder that consists of two round disks made of Nitinol wire mesh and linked together by a short connecting waist. Sixteen patients with secundum ASD met established two- and three-dimensional echocardiographic and cardiac catheterization criteria for transcatheter closure. The Amplatzer's size was chosen to be equal to or 1 mm less than the stretched diameter. The device was advanced transvenously into a 7F long guiding sheath and deployed under fluoroscopic and ultrasound guidance. Once its position was optimal, it was released.
The mean ASD diameter by transesophageal echocardiography was 14.1+/-2.3 mm and was significantly smaller (p < 0.001) than the stretched diameter of the ASD (16.8+/-2.4 mm). The mean device diameter was 16.6+/-2.3 mm. No complications were observed. After deployment of the prosthesis, there was no residual shunt in 13 (81.3%) of 16 patients. In three patients there was trivial residual shunt immediately after the procedure that had disappeared in two of them at the 3-month follow-up.
The Amplatzer is an efficient prosthesis that can be safely applied in children with secundum ASD. However, a study including a large number of patients and a longer follow-up period are required before this technique can be widely used.
本研究报告了我们使用新型封堵装置Amplatzer经导管闭合儿童继发孔型房间隔缺损(ASD)的临床经验。
以前用于经导管闭合房间隔交通的装置均未获得广泛认可。
我们检测了Amplatzer的有效性和安全性,它是一种新型的自定心房间隔封堵器,由两个由镍钛诺丝网制成的圆盘组成,并通过一个短的连接腰部连接在一起。16例继发孔型ASD患者符合经导管闭合的二维和三维超声心动图及心导管检查标准。选择的Amplatzer尺寸等于或比拉伸直径小1mm。该装置经静脉推进到一个7F长的引导鞘中,并在荧光透视和超声引导下展开。一旦其位置最佳,就将其释放。
经食管超声心动图测得的ASD平均直径为14.1±2.3mm,明显小于ASD的拉伸直径(16.8±2.4mm)(p<0.001)。装置平均直径为16.6±2.3mm。未观察到并发症。植入封堵器后,16例患者中有13例(81.3%)无残余分流。3例患者术后即刻有微量残余分流,其中2例在3个月随访时消失。
Amplatzer是一种有效的封堵器,可安全应用于继发孔型ASD患儿。然而,在该技术能够广泛应用之前,需要进行一项纳入大量患者且随访期更长的研究。