Morishita Y, Rubin S J, Hicks D G, Kwok E, Seo G S, Totterman S M
Department of Radiology, Toho University School of Medicine, Ohashi Hospital, Tokyo, Japan.
Acad Radiol. 1998 May;5(5):365-73. doi: 10.1016/s1076-6332(98)80155-4.
The authors' purpose was to develop a magnetic resonance (MR) imaging technique for examining the hip joint of a rabbit with a clinical MR imager.
Fourteen hips of 10 male New Zealand white rabbits were examined on a 1.5-T MR imager with a specially designed surface coil. Field of view was 3-8 cm, matrix was 256-512 x 192-256, section thickness was 0.9-2.0 mm, and spacing was 0.3-0.5 mm. The hips were sectioned and evaluated by means of light microscopy. Measurements of cartilage thickness from MR images were correlated with those from histologic specimens.
The resolution obtainable with a 1.5-T imager was adequate for imaging articular cartilage when a débrided rabbit hip specimen was placed in the center of a 3.5-cm single-loop coil. Rabbit hip cartilage had a trilaminar appearance on MR images. The coefficient of correlation between cartilage-thickness measurements was .81.
MR imaging of rabbit hip cartilage can be performed on 1.5-T clinical MR imaging unit. Limitations due to the field strength can be overcome with efficiently designed surface coils.
作者的目的是开发一种利用临床磁共振成像仪检查兔髋关节的磁共振成像(MR)技术。
使用专门设计的表面线圈,在1.5-T磁共振成像仪上对10只雄性新西兰白兔的14个髋关节进行检查。视野为3 - 8厘米,矩阵为256 - 512×192 - 256,层厚为0.9 - 2.0毫米,间距为0.3 - 0.5毫米。对髋关节进行切片并通过光学显微镜进行评估。将磁共振图像上软骨厚度的测量值与组织学标本的测量值进行相关性分析。
当将清创后的兔髋关节标本置于3.5厘米单环线圈中心时,1.5-T成像仪可获得的分辨率足以对关节软骨进行成像。兔髋关节软骨在磁共振图像上呈现出三层结构。软骨厚度测量值之间的相关系数为0.81。
兔髋关节软骨的磁共振成像可在1.5-T临床磁共振成像设备上进行。通过高效设计的表面线圈可以克服因场强导致的局限性。