Suppr超能文献

机构间床边血糖监测项目特征、准确性表现及质量控制文件的比较:美国病理学家学会Q-Probes对226家小型医院床边血糖监测的研究

Interinstitutional comparison of bedside blood glucose monitoring program characteristics, accuracy performance, and quality control documentation: a College of American Pathologists Q-Probes study of bedside blood glucose monitoring performed in 226 small hospitals.

作者信息

Novis D A, Jones B A

机构信息

Department of Pathology, Wentworth-Douglass Hospital, Dover, NH 03820, USA.

出版信息

Arch Pathol Lab Med. 1998 Jun;122(6):495-502.

PMID:9625416
Abstract

OBJECTIVES

To assess the accuracy of bedside blood glucose monitoring (BGM) in small hospitals, to assess the compliance with which hospital workers performing bedside BGM adhere to quality control (QC) procedures, and to identify those practice characteristics in small hospitals that are associated with better BGM accuracy and with better performance of BGM QC.

DESIGN

Over a 1-month period in 1996, voluntary participants in the College of American Pathologists Q-Probes laboratory quality improvement program prospectively compared glucose results of 30 split samples run on BGM instruments with those performed on laboratory glucose analyzers, collected quality control data on up to five inpatient BGM instruments, and completed questionnaires profiling BGM practice characteristics in their institutions.

SETTING AND PARTICIPANTS

Two hundred twenty-six hospitals with 200 or fewer occupied beds.

MAIN OUTCOME MEASURES

The percentages of glucose determinations performed on BGM instruments differing by more than 10%, 15%, and 20% from those split-sample results performed on laboratory glucose analyzers; the percent of BGM QC determinations required by institutions' BGM QC programs that BGM operators actually performed; and the percent of patient values reported when BGM QC was documented to be out of range and uncorrected, or reported when BGM QC was not performed at all.

RESULTS

Of 6095 split-specimen glucose results that participants simultaneously performed on BGM instruments and on laboratory glucose analyzers, 45.6% differed from each other by more than 10%, approximately 25% differed from each other by more than 15%, and almost 14% differed from each other by more than 20%. Of 216 laboratories that performed at least 30 QC events during the study period, slightly over a third completed 100% of their required QC determinations, and 10% completed, at most, 77% of their required BGM QC determinations. Of 115,973 BGM determinations that participants reported on hospitalized patients, 3.3% were reported when QC was either out of range or when there was no documentation that QC had been performed at all. Better accuracy and/or better QC performance was associated with laboratory personnel rather than nursing personnel both supervising institutions' BGM QC programs and running institutions' daily routine BGM QC; with BGM operators both routinely running three, rather than two, levels of QC analytes; with BGM operators regularly comparing BGM results with laboratory analyzer glucose results; and with institutions participating in external proficiency programs. Institutions that completed all required BGM QC tasks tended to perform better on the BGM accuracy study than did those institutions that completed, at most, 77% of their required QC.

CONCLUSIONS

We found the rates of BGM accuracy and of QC performance adequacy achieved in small hospitals to be similar to those determined in previous Q-Probes studies conducted in large institutions. A significant amount of institutional bedside testing does not meet current standards for accuracy or for quality control. Some institutions may improve their accuracy and/or QC performances by having laboratory personnel intimately involved in their institution's BGM QC program, by routinely comparing BGM results with those performed using glucose analyzers in the clinical laboratory, by routinely running three rather than two glucose QC control levels, by participating in external proficiency programs, and by strictly adhering to institutional QC protocols.

摘要

目的

评估小型医院床旁血糖监测(BGM)的准确性,评估医院工作人员进行床旁BGM时遵循质量控制(QC)程序的情况,并确定小型医院中与更好的BGM准确性和更好的BGM QC表现相关的实践特征。

设计

1996年的1个月期间,美国病理学家学会Q-Probes实验室质量改进计划的自愿参与者前瞻性地比较了在BGM仪器上检测的30个分离样本的血糖结果与在实验室血糖分析仪上检测的结果,收集了多达五台住院患者BGM仪器的质量控制数据,并完成了有关其机构中BGM实践特征的问卷调查。

设置与参与者

226家床位不超过200张的医院。

主要观察指标

在BGM仪器上进行的血糖测定结果与在实验室血糖分析仪上进行的分离样本结果相比,相差超过10%、15%和20%的百分比;机构BGM QC计划要求的BGM QC测定中BGM操作人员实际执行的百分比;以及当记录的BGM QC超出范围且未校正时报告的患者值百分比,或当根本未进行BGM QC时报告的百分比。

结果

参与者在BGM仪器和实验室血糖分析仪上同时检测的6095个分离样本血糖结果中,45.6%彼此相差超过10%,约25%彼此相差超过15%,近14%彼此相差超过20%。在研究期间至少进行30次QC事件的216个实验室中,略超过三分之一完成了100%的所需QC测定,10%最多完成了77%的所需BGM QC测定。在参与者报告的115973例住院患者BGM测定中,3.3%是在QC超出范围或根本没有记录QC已执行的情况下报告的。更好的准确性和/或更好的QC表现与监督机构BGM QC计划并运行机构日常常规BGM QC的实验室人员而非护理人员有关;与BGM操作人员常规运行三个而非两个水平的QC分析物有关;与BGM操作人员定期将BGM结果与实验室分析仪血糖结果进行比较有关;以及与参与外部能力验证计划的机构有关。完成所有所需BGM QC任务的机构在BGM准确性研究中的表现往往优于最多完成77%所需QC的机构。

结论

我们发现小型医院中BGM准确性和QC表现充分性的比率与先前在大型机构中进行的Q-Probes研究确定的比率相似。大量的机构床旁检测不符合当前的准确性或质量控制标准。一些机构可以通过让实验室人员密切参与其机构的BGM QC计划、定期将BGM结果与临床实验室中使用血糖分析仪获得的结果进行比较、常规运行三个而非两个血糖QC控制水平、参与外部能力验证计划以及严格遵守机构QC方案来提高其准确性和/或QC表现。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验