Clinical in vivo dosimetry using optical fibers.

作者信息

Gripp S, Haesing F W, Bueker H, Schmitt G

机构信息

Department of Radiation Oncology, Heinrich Heine University, Duesseldorf, Germany.

出版信息

Radiat Oncol Investig. 1998;6(3):142-9. doi: 10.1002/(SICI)1520-6823(1998)6:3<142::AID-ROI5>3.0.CO;2-Y.

Abstract

Discoloring of glass due to ionizing radiation depends on the absorbed dose. The radiation-induced light attenuation in optical fibers may be used as a measure of the dose. In high-energy photon beams (6 MV X rays), a lead-doped silica fiber can be calibrated. A dosimeter based on an optical fiber was developed for applications in radiation therapy. The diameter of the mounted fiber is 0.25 mm, whereas the length depends on the sensitivity required. To demonstrate the applicability, a customized fiber device was used to determine scattered radiation close to the lens of the eye. Measurements were compared with TLDs (LiF) in an anthropomorphic phantom. The comparison with TLD measurements shows good agreement. In contrast to TLD, optical fibers provide immediate dose values, and the readout procedure is much easier. Owing to its small size and diameter, interesting invasive dose measurements are feasible.