Song J Z, Chen H F, Tian S J, Sun Z P
Division of Instrumental Analysis, National Institute for the Control of Pharmaceutical and Biological Products, Temple of Heaven, Beijing, China.
J Chromatogr B Biomed Sci Appl. 1998 Apr 24;708(1-2):277-83. doi: 10.1016/s0378-4347(97)00635-x.
A capillary electrophoresis method was described for the determination of metformin in human plasma based on the extraction of the ion-pair with bromothymol blue into chloroform. Phenformin was used as internal standard. Field-amplified sample stacking injection was employed with an electrokinetic injection voltage of 10 kV for 10 s. The running buffer was 0.1 M phosphate buffer (pH 2.5), running voltage was 20 kV and the UV absorbance detection was set at 195 nm. The limit of quantitation was 0.25 microg/ml. Linearity range of calibration curve was 0.25 to 3.5 microg/ml. Recoveries for three levels (0.25, 1 and 2 microg/ml) were 80.24%, 67.44% and 58.97% (n = 5 for each level), respectively. The intra-day precisions for the three levels were 11.9%, 3.09% and 4.33% and the inter-day precisions were 12.4%, 4.57% and 4.94%, respectively. The concentrations of metformin hydrochloride in human plasma of eight volunteers were measured after orally administrating metformin enteric-capsule and tablet.