超滤对中空纤维人工肾溶质清除率的影响。

Effects of ultrafiltration on solute clearances in hollow fiber artificial kidneys.

作者信息

Nolph K D, New D L

出版信息

J Lab Clin Med. 1976 Oct;88(4):593-600.

PMID:965810
Abstract

Although solute clearances in artificial kidney coils increase with ultrafiltration (UF), we have previously shown that increases are usually less than UF rate (most likely because of decreases in diffusive transport with UF coils and, for larger solutes, molecular sieving). The present studies demonstrate the effects of UF on clearances of Na and bromsulphalein (BSP) (mol. wt. 838) in hollow fiber dialyzers. Clearances were measured at increasing transmembrane hydrostatic pressures at perfusion rates of 200 and 500 ml. per minute. Fractions of total clearance attributable to diffusion as compared to solvent drag forces were calculated. Sieving coefficients were determined in studies where diffusion was minimized and clearance was primarily by solvent drag. Clearance increases were less than UF rate only for BSP; molecular sieving most likely accounts for the difference at high perfusion rates. Only at 200 ml. per minute was slight decrease of diffusion with UF suggested. Thus, in contrast to coils, there is minimal or no decrease in diffusion with UF in hollow fiber dialyzers.

摘要

尽管人工肾盘管中的溶质清除率会随着超滤(UF)增加,但我们之前已经表明,清除率的增加通常小于超滤速率(很可能是因为超滤盘管中扩散传输的减少,以及对于较大溶质而言的分子筛分作用)。本研究证明了超滤对中空纤维透析器中钠和溴磺酚酞(BSP,分子量838)清除率的影响。在每分钟200和500毫升的灌注速率下,随着跨膜静水压力的增加测量清除率。计算了与溶剂拖曳力相比,归因于扩散的总清除率的分数。在扩散最小化且清除主要通过溶剂拖曳的研究中确定了筛分系数。仅对于BSP,清除率的增加小于超滤速率;分子筛分很可能是高灌注速率下差异的原因。仅在每分钟200毫升时,提示超滤会使扩散略有降低。因此,与盘管相反,中空纤维透析器中超滤时扩散的降低最小或没有降低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验