A quantitative rotational test of vestibular function.

作者信息

McClure J A, Lycett P, Bicker G R

出版信息

J Otolaryngol. 1976 Aug;5(4):279-88.

PMID:966328
Abstract

This paper describes a quantitative vestibular test in which the stimulus is a constant angular acceleration of 3 degrees/sec2 for 80 sec. The nystagmus output is plotted as cumulative slow phase eye displacement against time and is displayed on an oscilloscope screen along with the output from a mathematical model describing vestibulo-oculomotor function. External dials allow one to change the equation constants thereby altering the shape of the model output. Values for equation constants which match the model output to the patient can be read directly from the dials, thus providing a description of the vestibular system under test. Results indicate that vestibular asymmetry (manifest by spontaneous nystagmus is not uncommon in normals, and only when spontaneous nystagmus is coupled with vestibular gain does one get a measure of abnormality. The significance of altered vestibular parameters in certain pathological states is discussed.

摘要

本文描述了一种定量前庭测试,其中刺激是80秒内3度/秒²的恒定角加速度。眼震输出表示为累积慢相眼位移随时间的变化,并与描述前庭眼动功能的数学模型的输出一起显示在示波器屏幕上。外部刻度盘允许人们改变方程常数,从而改变模型输出形状。使模型输出与患者相匹配的方程常数的值可直接从刻度盘上读出,从而提供对受试前庭系统的描述。结果表明,前庭不对称(表现为正常人中并不罕见的自发性眼震),只有当自发性眼震与前庭增益相结合时,才会得到异常的测量值。文中讨论了某些病理状态下前庭参数改变的意义。