Lee W N, Lim S, Bassilian S, Bergner E A
Department of Pediatrics, Harbor-UCLA Medical Center, Torrance 90502, USA.
J Mass Spectrom. 1998 Jul;33(7):627-30. doi: 10.1002/(SICI)1096-9888(199807)33:7<627::AID-JMS677>3.0.CO;2-7.
A method is described for the calibration of an isotope ratio mass spectrometry working standard for 2H/1H analysis using highly purified (> 99.99% enriched) deuterium oxide (2H2O). Serial dilution of such a compound can be used to construct a standard curve for calibration purposes using the relationship [formula: see text] where APE is the atom percent excess. It was found that the determined rWR had a precision of +/- 1%. Most of the variation came from sample handling, which includes the serial dilution of the weighed deuterium oxide, the reduction of water to hydrogen gas and the introduction of the hydrogen gas into the mass spectrometer. The use of this calibration procedure allows the direct determination of the isotope ratio rSA, and APE from a standard curve over a wide range of enrichment. Unlike the reference standards V-SMOW and SLAP, the highly purified deuterium oxide is non-exhaustible and readily available. The highly purified deuterium oxide should be used in place of secondary standards to calibrate working references for 2H/1H analysis in biological studies.
描述了一种使用高纯度(富集度>99.99%)氧化氘(2H2O)对用于2H/1H分析的同位素比率质谱工作标准进行校准的方法。这种化合物的系列稀释可用于构建校准标准曲线,利用公式[公式:见原文],其中APE是原子百分超量。发现所测定的rWR精度为±1%。大部分变化来自样品处理,这包括对称重的氧化氘进行系列稀释、将水还原为氢气以及将氢气引入质谱仪。使用这种校准程序可在广泛的富集范围内从标准曲线直接测定同位素比率rSA和APE。与参考标准V-SMOW和SLAP不同,高纯度氧化氘是不可耗尽且易于获得的。在生物学研究中,应使用高纯度氧化氘代替二级标准来校准用于2H/1H分析的工作参考标准。