Suppr超能文献

通过相位对比磁共振血管造影(PC-MRA)与双功超声检查对比进行血液透析分流处血流定量分析

[Blood flow quantification in hemodialysis shunts by phase contrast magnetic resonance angiography (PC-MRA) compared with duplex sonography].

作者信息

Gutberlet M, Venz S, Kahl A, Ehrenstein T, Puls R, Hosten N, Frei U, Felix R

机构信息

Strahlenklinik und Poliklinik, Charité, Medizinische Fakultät der Humboldt-Universität Berlin.

出版信息

Rofo. 1998 Aug;169(2):163-9. doi: 10.1055/s-2007-1015068.

Abstract

PURPOSE

Aim of this study was to evaluate whether phase contrast MR angiography (PC-MRA) could provide additional functional information besides morphology in the assessment of haemodialysis fistulae.

MATERIAL AND METHOD

Twenty-two patients (11 male, 11 female), aged 22-77 years, were examined. MR images were obtained with a 1.5 T Gyroscan ACS-NT (Philips, Best, Netherlands) using a high-resolution wrap-around coil. In addition to MRA blood-flow measurements were performed with a gradient-echo sequence (TR 14 ms, TE 5-5.5 ms, flip-angle 15 degrees, 6 mm slice thickness, retrospective gating, matrix 96:128) in the venous and arterial section of the fistulae. Doppler flow measurements were performed at the same position with a Sonoline Elegra (Siemens AG, Erlangen) using a 7.5 MHz transducer.

RESULTS

Both methods of flow-volume measurements showed a good correlation (r = 0.94 in the arterial section, r = 0.90 in the venous section, p < 0.001). The average calculated blood flow was measured 11% (arterial section) and 12.8% (venous section) higher with Pulsed Waved Doppler as compared to PC-MRA.

CONCLUSION

PC-MRA with a high-resolution wrap-around coil is a reliable method for measuring functional parameters like flow-volume and flow velocity in haemodialysis fistulae and a useful complement to the visualisation capabilities of MRA.

摘要

目的

本研究的目的是评估相位对比磁共振血管造影(PC-MRA)在评估血液透析瘘管时,除形态学外是否能提供额外的功能信息。

材料与方法

对22例患者(11例男性,11例女性)进行检查,年龄在22 - 77岁之间。使用1.5T Gyroscan ACS-NT(飞利浦,贝斯特,荷兰)磁共振成像仪,采用高分辨率环绕线圈获取图像。除MRA外,还在瘘管的静脉和动脉段使用梯度回波序列(TR 14ms,TE 5 - 5.5ms,翻转角15度,层厚6mm,回顾性门控,矩阵96:128)进行血流测量。使用配备7.5MHz探头的Sonoline Elegra(西门子股份公司,埃尔朗根)在相同位置进行多普勒血流测量。

结果

两种血流量测量方法显示出良好的相关性(动脉段r = 0.94,静脉段r = 0.90,p < 0.001)。与PC-MRA相比,脉冲波多普勒测量的平均计算血流量在动脉段高11%,在静脉段高12.8%。

结论

使用高分辨率环绕线圈的PC-MRA是一种可靠的方法,可用于测量血液透析瘘管的血流量和流速等功能参数,是MRA可视化能力的有益补充。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验