[人主动脉瓣与肺动脉瓣力学性能的比较]

[Comparison of mechanical properties between human aortic valve and pulmonary arterial valve].

作者信息

Li Z, Cheng M, Yu J, Cheng J

机构信息

Dept. of Civil Engineering & Applied Mechanics, Sichuan Union University, Chengdu.

出版信息

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 1997 Jun;14(2):115-7.

PMID:9817637
Abstract

We have investigated the mechanical properties of Chinese aortic and pulmonary valves by means of uniaxial tension. The parameters include the stress-strain curve, the maximum tangent modulus E (the slope of linear segment) in the curve, the tensile strength delta b, and the ultimate strain epsilon b. The results have shown that E, delta b, and epsilon b values between aortic valve and pulmonary valve exist significant differences (significant level = 0.01), but the mechanical property differences between circumferential direction and radial direction in the pulmonary arterial valve are not significant, that means, the valve appears small anisotropic behaviour.

摘要

我们通过单轴拉伸研究了中国主动脉瓣和肺动脉瓣的力学性能。参数包括应力-应变曲线、曲线中的最大切线模量E(线性段的斜率)、拉伸强度δb和极限应变εb。结果表明,主动脉瓣和肺动脉瓣之间的E、δb和εb值存在显著差异(显著水平=0.01),但肺动脉瓣周向和径向之间的力学性能差异不显著,也就是说,瓣膜表现出较小的各向异性行为。