Cerclage wiring technique after proximal femoral fracture in total hip arthroplasty.

作者信息

Fishkin Z, Han S M, Ziv I

机构信息

Department of Orthopaedic Surgery, State University of New York at Buffalo, USA.

出版信息

J Arthroplasty. 1999 Jan;14(1):98-101. doi: 10.1016/s0883-5403(99)90209-7.

Abstract

Cerclage wires have been used to stabilize proximal femoral cracks after stem insertion in cementless total hip arthroplasty. The objective of this study was to determine the optimal number and orientation of cerclage wires necessary to prevent stem subsidence and crack propagation. The crack was stabilized by 1, 2, or 3 wires placed either normal to the femoral neck axis or normal to the crack. The femora were compressed to 2,670 N while measuring crack opening and stem subsidence. Wires placed normal to the crack allowed less stem subsidence by 3.17 mm and less crack opening by 1.55 mm compared with wires placed normal to the neck. The addition of multiple wires reduced subsidence by 50% and reduced crack opening to <1 mm. Medial and anterior calcar cracks are best stabilized by at least 2 cerclage wires that are placed normal to the crack.