Coleman R R, Bernard B B, Harrison D E, Harrison S O
Chiropractic Biophysics Non-profit Inc, Harvest, Alabama, USA.
J Manipulative Physiol Ther. 1999 Jan;22(1):21-5. doi: 10.1016/s0161-4754(99)70101-6.
Historically, measurement of 2-dimensional (2-D) radiographic images on the anteroposterior radiograph has been made to assess 3-dimensional (3-D) y-axis vertebral rotations.
To correlate and quantify measurements of the projected 2-D radiographic image with the degree of 3-D y-axis rotation.
A computer model was positioned in a simulated x-ray beam. Points of model contact with the simulated beam were projected onto a line in the neutral position and the first 7 degrees of both positive and negative y-axis rotation using two different axes of rotation. A larger model, a shape-altered model, and a decreased source-object-distance model were also studied.
3-D y-axis rotation of vertebrae causes an off-center displacement of the 2-D projected lamina junction in relation to the projected vertebral body. The magnitude of displacement increases with increasing degrees of rotation. In our model, no clinically significant difference was found in the amount of the projected off-center displacement of the lamina junction between either of our two chosen axes of rotation. However, significant differences in the projected offset were found between vertebrae with the same degree of rotation as a result of changes in vertebral shape, size, and positioning. The projected lamina junction off-centering at a given rotation is quantified for our model.
Use of millimetric measurement of the projected lamina offset on the anteroposterior radiograph is an inaccurate method for the assessment of the degree of 3-D y-axis vertebral rotation.
从历史上看,一直通过测量前后位X线片上的二维(2-D)放射影像来评估三维(3-D)y轴椎体旋转情况。
将二维放射影像投影测量值与三维y轴旋转程度进行关联并量化。
将一个计算机模型置于模拟X线束中。模型与模拟线束的接触点投影到一条处于中立位置的线上,以及使用两个不同旋转轴在正负y轴旋转的前7度范围内进行投影。还研究了一个更大的模型、一个形状改变的模型和一个源-物体距离减小的模型。
椎体的三维y轴旋转会导致二维投影的椎板交界处相对于投影椎体发生偏心移位。移位幅度随旋转度数增加而增大。在我们的模型中,在我们选择的两个旋转轴中,椎板交界处投影偏心移位量在临床上未发现显著差异。然而,由于椎体形状、大小和位置的变化,在相同旋转度数的椎体之间发现投影偏移存在显著差异。对我们的模型而言,在给定旋转角度下投影的椎板交界处偏心情况得以量化。
通过测量前后位X线片上投影的椎板偏移量(以毫米计)来评估三维y轴椎体旋转程度是一种不准确的方法。