Evaluation of a laser Doppler flowmeter to assess blood flow in human primary incisor teeth.

作者信息

Fratkin R D, Kenny D J, Johnston D H

机构信息

Department of Dentistry, Hospital for Sick Children, Toronto, Canada.

出版信息

Pediatr Dent. 1999 Jan-Feb;21(1):53-6.

Abstract

PURPOSE

This study determined whether the portable Advance Laser Flowmeter Model 21, a clinical instrument, would indicate significant values when pulpal blood flow was present or absent; whether the analog output of the flowmeter was time-linked to the heart rate; and whether labial and lingual crown surfaces produced different flow values.

METHODS

Teeth were tested under two conditions: with the teeth in situ and extracted, or with the pulp present and removed.

RESULTS

Values before and after the teeth were extracted, and those before and after pulpectomy with the tooth in situ were significantly different. These findings verified that the flowmeter measured the presence of bloodflow. Peaks of the electrocardiogram waves and the regular signal fluctuations of the flowmeter were time-linked and verified the measurement of the heart rate from the dental pulp. Differences in the values for the lingual and labial surfaces of individual teeth were not significant.

CONCLUSIONS

This instrument proved a valid means of determining the presence of pulpal blood flow in primary incisors.

摘要

目的

本研究旨在确定临床仪器便携式21型高级激光流量计在牙髓血流存在或不存在时是否会显示出显著数值;流量计的模拟输出是否与心率存在时间关联;以及唇面和舌面牙冠表面是否会产生不同的流量值。

方法

在两种情况下对牙齿进行测试:牙齿在原位和拔除后,或牙髓存在和去除后。

结果

牙齿拔除前后以及原位牙髓摘除前后的值存在显著差异。这些发现证实了流量计能够测量血流的存在。心电图波峰与流量计的规则信号波动存在时间关联,并证实了从牙髓测量心率。单个牙齿唇面和舌面的值差异不显著。

结论

该仪器被证明是确定乳切牙髓血流存在的有效手段。