Venous saturation monitoring: reliability of the Bentley SM-0200 OxySAT with and without use of a cuvette.

作者信息

Bellaiche A L, Christensen T D

机构信息

Department of Cardio-thoracic and Vascular Surgery, Aarhus University Hospital, Skejby Hospital, Denmark.

出版信息

Perfusion. 1999 Jan;14(1):69-76. doi: 10.1177/026765919901400110.

Abstract

In-line venous saturation monitoring is a useful method of quality management during extracorporeal circulation (ECC). The Bentley SM-0200 OxySAT meter was tested with and without the use of the recommended cuvette. The study consisted of an in vitro part, where extremely low saturations could be safely measured, and an in vivo part, carried out under routine ECC conditions. The results show that the OxySAT may be used with or without a cuvette, and that it needs calibration for optimal accuracy in either case.

摘要

在线静脉饱和度监测是体外循环(ECC)期间质量管理的一种有用方法。对Bentley SM - 0200血氧饱和度仪在使用和不使用推荐比色皿的情况下进行了测试。该研究包括一个体外部分,在此部分可以安全地测量极低的饱和度,以及一个体内部分,在常规体外循环条件下进行。结果表明,血氧饱和度仪在使用或不使用比色皿的情况下均可使用,并且在这两种情况下都需要进行校准以获得最佳准确性。