Bassi A S, Tang D, Bergougnou M A
Department of Chemical and Biochemical Engineering, Faculty of Engineering Science, University of Western Ontario, London, Ontario, N6A 5B9,
Anal Biochem. 1999 Mar 15;268(2):223-8. doi: 10.1006/abio.1998.3082.
In this study, an amperometric carbon paste biosensor is developed for glucose-6-phosphate (G6P) monitoring which is based on entrapped Mg2+ ions, G6P dehydrogenase, NADP+ polyethylenimine (PEI) and the electroactive mediator, tetracyanoquinodimethane (TCNQ). The calibration line had a slope of 1.55 x 10(-5) A. M-1 with a correlation coefficient of 0.9965. The limit of detection (defined as three times the standard deviation of the response of the electrode to blank phosphate buffer injections (noise)) of the G6P biosensor was 5.0 x 10(-5) M. The application of this biosensor for monitoring G6P in human blood using the standard addition method is also demonstrated. A two-parameter empirical equation which adequately describes the deactivation of the biosensor steady-state response with time is also proposed.
在本研究中,开发了一种用于监测6-磷酸葡萄糖(G6P)的安培型碳糊生物传感器,其基于包埋的Mg2+离子、G6P脱氢酶、NADP+、聚乙烯亚胺(PEI)和电活性介质四氰基对苯二醌二甲烷(TCNQ)。校准曲线的斜率为1.55×10(-5) A·M-1,相关系数为0.9965。G6P生物传感器的检测限(定义为电极对空白磷酸盐缓冲液注射响应的标准偏差的三倍(噪声))为5.0×10(-5) M。还展示了使用标准加入法将该生物传感器应用于监测人体血液中的G6P。此外,还提出了一个能充分描述生物传感器稳态响应随时间失活的双参数经验方程。