Joglar J A, Hamdan M H, Welch P J, Page R L
Department of Internal Medicine, The University of Texas Southwestern Medical Center, Dallas 75235-9047, USA.
Am J Cardiol. 1999 Mar 1;83(5):790-2, A10. doi: 10.1016/s0002-9149(98)00994-1.
Dry-electrode heart rate monitors allow display of heart rate by transmitting a signal to the receiving device, which typically is on the wrist or exercise machine, but due to the potential for electromagnetic interference, their use has been contraindicated in patients with pacemakers. In 12 patients, we found no adverse effect on pacemaker function; in addition, the monitors generally were accurate in measuring heart rate during pacing.
干电极心率监测仪通过向接收设备传输信号来显示心率,接收设备通常戴在手腕上或安装在健身器材上,但由于存在电磁干扰的可能性,起搏器患者禁用此类监测仪。我们对12名患者进行了研究,发现该监测仪对起搏器功能无不良影响;此外,在起搏过程中,该监测仪通常能准确测量心率。