High-performance liquid chromatography column switching applied to the trace determination of herbicides in environmental and drinking water samples.

作者信息

Nemth Konda L, Morovján G, Csokan P

机构信息

Analytical Chemsitry Department, State Control Institute for Veterinary Biologicals, Drugs and Feeds, Budapest, Hungary.

出版信息

J Chromatogr Sci. 1999 Mar;37(3):71-6. doi: 10.1093/chromsci/37.3.71.

Abstract

A selective and sensitive coupled-column high-performance liquid chromatographic method is developed for the simultaneous determination of 5 phenylurea herbicides (monuron, linuron, isoproturon, monolinuron, and diuron) in environmental and drinking water samples. Sample clean-up is performed automatically by means of a column switching technique. Using 2 octadecyl silica columns connected via two programmable 6-port valves and ultraviolet detection at 244 nm, the aforementioned compounds can be determined at the low concentration levels required for pesticide residue analysis in water samples. A mobile phase consisting of a mixture of methanol-water (55:45, v/v) is pumped at 1 mL/min. For the 5 phenylureas, high recoveries ranging from 94.9 to 101.6%, good reproducibility with relative standard deviations lower than 5%, and wide linear ranges up to 20 micrograms/L are observed with determination limits of 0.05 microgram/L. The method is successfully applied to the screening of different environmental water samples such as surface, ground, rain, and drinking water.