Pandian N G, Schwartz S L, Hsu T L, Weintraub A R, Katz S E, Aronovitz M, Konstam M A, Salem D N, Kreis A
Departments of Medicine and Radiology, New England Medical Center Tufts University School of Medicine, Boston, MA 02111.
Echocardiography. 1991 Jan;8(1):127-34. doi: 10.1111/j.1540-8175.1991.tb01410.x.
Recently catheter-based ultrasound devices have become available for obtaining high-resolution images of blood vessels. In this study we evaluated the feasibility of imaging cardiac structures using 20-MHz ultrasound catheters. In 25 dogs, the ultrasound catheter was advanced into the right and left heart chambers percutaneously. The intravascular devices yielded images of the right atrial wall, right and left ventricular myocardia, tricuspid, pulmonic, and aortic valves, and the great vessels. Although the small depth of field inherent to the frequency range of 20 MHz limited the visualization to only portions of the cardiac chambers, the images obtained were of high resolution and allowed easy identification of the various cardiac structures. Intracardiac echocardiography was easy to perform and did not result in damage to the cardiac structures. We conclude that intracardiac echocardiography using ultrasound catheters provides a new approach to cardiac imaging and that the development of lower frequency catheters could aid in extending the potential utility of intracardiac echocardiography.
最近,基于导管的超声设备已可用于获取血管的高分辨率图像。在本研究中,我们评估了使用20兆赫超声导管对心脏结构进行成像的可行性。在25只犬中,将超声导管经皮推进到右心腔和左心腔。血管内设备获取了右心房壁、右心室和左心室心肌、三尖瓣、肺动脉瓣和主动脉瓣以及大血管的图像。尽管20兆赫频率范围内固有的小景深限制了仅对心腔部分的可视化,但所获得的图像具有高分辨率,并且能够轻松识别各种心脏结构。心内超声心动图操作简便,不会对心脏结构造成损伤。我们得出结论,使用超声导管的心内超声心动图为心脏成像提供了一种新方法,并且开发低频导管可能有助于扩大心内超声心动图的潜在应用。