[The noninvasive determination of cardiac output by means of impedance cardiography. Comparative evaluation with a thermal dilution technique (author's transl)].

作者信息

Nechwatal W, Bier P, Eversmann A, König E

出版信息

Basic Res Cardiol. 1976 Sep-Oct;71(5):542-52. doi: 10.1007/BF01909768.

Abstract

When a high frequency alternating current is passed through the thorax by means of tape-electrodes, the electrical impedance Z is changing synchronously with the cardiac cycle. The relation between the first derivative dZ/dt and the stroke volume is the basis for the application of impedance cardiography (IK) Determinations of cardiac output (CO) at rest during ergometer exercise using this noninvasive technique were compared with CO values obtained by the thermal dilution method. In 57 subjects simultaneous determinations using both methods were performed at resting conditions. The correlation coefficient was r = 0.63. In addition, in 48 of the subjects CO was measured with both methods during supine ergometer exercise. Comparison of the exercise values resulted in a correlation coefficient of r = 0.63. In all 48 cases there was a concordant change of the IK and thermodilution CO values. It is concluded that relative changes of CO are reliably detected by IK. The absolute values can be utilized only approximately.

摘要

当高频交流电通过带状电极穿过胸部时,电阻抗Z随心动周期同步变化。一阶导数dZ/dt与每搏输出量之间的关系是阻抗心动图(IK)应用的基础。使用这种非侵入性技术,对测力计运动期间静息状态下的心输出量(CO)测定结果与通过热稀释法获得的CO值进行了比较。在57名受试者中,在静息状态下同时使用两种方法进行了测定。相关系数为r = 0.63。此外,在48名受试者中,在仰卧测力计运动期间用两种方法测量了CO。对运动值的比较得出相关系数为r = 0.63。在所有48例中,IK和热稀释CO值均有一致变化。得出的结论是,IK能够可靠地检测到CO的相对变化。绝对值只能大致使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验