Autoclaving impairs the connector-tube bond of the laryngeal mask airway but not its airtightness.

作者信息

Preis C, Hartmann T, Preis I, Wildling E, Gilly H

机构信息

Department of Anaesthesiology and General Intensive Care, University of Vienna, Austria.

出版信息

Br J Anaesth. 1998 Nov;81(5):795-6. doi: 10.1093/bja/81.5.795.

Abstract

The general-purpose laryngeal mask airway (LMA) is re-usable when undamaged, and cleaned and autoclaved correctly. We had found weakening of the silicone adhesive that bonds the connector of the LMA to the tube. We report that repeated autoclaving damaged the adhesive such that the connector could be rotated in the tube after the 12th autoclave cycle in almost all of the LMA tested. The damage to the adhesive did not affect the airtightness of the junction, which appears to be maintained by the material properties of the connector and tube and by the shape of the join.