Suppr超能文献

轻度、中度和重度未经治疗的黑人高血压患者中重复自动测量与日间动态血压测量的比较

Repeated automated versus daytime ambulatory blood pressure measurement in mild, moderate and severe untreated black hypertensive patients.

作者信息

Mueller UK, Wells M, Radevski I, Ouwerkerk J, Tager R, Sliwa K, Sareli P

机构信息

Division of Cardiology, Baragwanath Hospital, University of the Witwatersrand, Johannesburg, South Africa.

出版信息

Blood Press Monit. 1997 Dec;2(1):21-25.

Abstract

OBJECTIVE

To examine the extent to which one session of repeated automated (30-45 min) blood pressure measurements in non-treated, hypertensive patients can be used to predict daytime ambulatory blood pressure (ABP). METHODS: Two hundred untreated black hypertensive patients (mean age 50.2+/-11.2) were retrospectively stratified on the basis of the daytime ambulatory diastolic blood pressure (DBP) into mild, moderate and severe hypertensives (90 mmHg <DBP < 140 mmHg). Office blood pressure was determined at 3 min intervals for a total of 30-45 min using the Dinamap Critikon 1846 SX monitor. Twenty-four hour ambulatory blood pressure (ABP) was performed on the same day in all patients using a SpaceLabs 90207 device. RESULTS: The mean diastolic Dinamap measurements compared with the mean ambulatory daytime DBP were very similar in terms of 'traditional' statistics (mean: 114 +/- 14.6 versus 113 +/- 12 mmHg, correlation: 0.80). However, agreement analysis according to Bland and Altman showed a poor agtreement with an error as high as 19 mmHg. The mean systolic Dinamap measurements compared with the mean ambulatory daytime SBP were significantly overestimated (mean: 185 +/- 25.0 versus 170 +/- 19.3 mmHg, correlation 0.82) and the error was as high as 43 mmHg. On residual analysis, 75% of the diastolic Dinamap values and only 35% of the systolic Dinamap measurements were within an error of +/- 10 mmHg. The average of the first five Dinamap measurements had the greatest predictive ability for the daytime ABP. CONCLUSION: As a result of the poor agreement between the two methods, we suggest that, from the clinical point of view, the two techniques should be treated as two separate entities.

摘要

目的

探讨在未经治疗的高血压患者中,进行一次重复自动(30 - 45分钟)血压测量在多大程度上可用于预测日间动态血压(ABP)。方法:200例未经治疗的黑人高血压患者(平均年龄50.2±11.2岁)根据日间动态舒张压(DBP)回顾性分层为轻度、中度和重度高血压患者(90 mmHg < DBP < 140 mmHg)。使用Dinamap Critikon 1846 SX监护仪每隔3分钟测定一次诊室血压,共测量30 - 45分钟。所有患者于同日使用SpaceLabs 90207设备进行24小时动态血压(ABP)监测。结果:就“传统”统计数据而言,Dinamap测量的平均舒张压与动态日间平均DBP非常相似(均值:114±14.6对113±12 mmHg,相关性:0.80)。然而,根据Bland和Altman法进行的一致性分析显示一致性较差,误差高达19 mmHg。Dinamap测量的平均收缩压与动态日间平均SBP相比显著高估(均值:185±25.0对170±19.3 mmHg,相关性0.82),误差高达43 mmHg。经残差分析,75%的Dinamap舒张压值和仅35%的Dinamap收缩压测量值误差在±10 mmHg以内。Dinamap前五次测量的平均值对日间ABP的预测能力最强。结论:由于两种方法之间一致性较差,我们建议,从临床角度来看,这两种技术应被视为两个独立的实体。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验