Intermittent P wave sensing in a patient with DDD pacemaker.

作者信息

Paraskevaidis S, Mochlas S, Hadjimiltiadis S, Louridas G

机构信息

First Cardiology Clinic, Aristotelian University, AHEPA Hospital, Thessaloniki, Greece.

出版信息

Pacing Clin Electrophysiol. 1999 Apr;22(4 Pt 1):689-90. doi: 10.1111/j.1540-8159.1999.tb00517.x.

Abstract

We present the case of a patient with a DDD pacemaker and intermittent P wave sensing due to T wave oversensing by the ventricular lead. The T wave sensing caused initiation of an extended atrial refractory period and the P waves, falling within this period, were not sensed. The problem was solved by decreasing the ventricular sensitivity.