Shandling A H, Castellanet M J, Messenger J C, Brownlee R R
Division of Cardiology, Memorial Medical Center of Long Beach, CA 90801-1428.
Pacing Clin Electrophysiol. 1988 Oct;11(10):1419-25. doi: 10.1111/j.1540-8159.1988.tb04990.x.
A 73-year-old male required a permanent pacemaker for complete heart block. After placement of the dual-chamber pacemaker, the patient developed random, repetitive ventricularly paced coupling at the upper rate limit. The diagnosis made from the surface electrocardiography was one of ventricular tracking of occult (T wave obscured) atrial ectopic activity. Examination of the telemetered atrial electrogram (EGM) revealed a diagnosis of intermittent far-field atrial sensing of the terminal ventricular depolarization. The value of obtaining an atrial electrogram during dual-chamber pacing is discussed.
一名73岁男性因完全性心脏传导阻滞需要植入永久性起搏器。植入双腔起搏器后,患者在最高心率限制时出现随机、重复性心室起搏耦联。体表心电图诊断为隐匿性(T波模糊)房性异位活动的心室跟踪。对遥测心房电图(EGM)的检查显示诊断为间歇性远场心房感知终末心室去极化。本文讨论了在双腔起搏期间获取心房电图的价值。