Evaluation of therapeutic ultrasound devices: performance stability over 44 weeks of clinical use.

作者信息

Rivest M, Quirion-de Girardi C, Seaborne D, Lambert J

出版信息

Physiother Can. 1987 Mar-Apr;39(2):77-86.

Abstract

The aim of this study is to establish at what time intervals regular checks of therapeutic ultrasound devices (TUSD) should be made to verify power output, timer accuracy, and ultrasonic frequency. Twenty-nine TUSDs from physiotherapy departments of 11 hospitals affiliated with the University of Montreal were retained for the study. The make, model, date of purchase, and other pertinent data of the TUSDs were determined from a questionnaire completed by each participating hospital. Power output, timer accuracy, and ultrasonic frequency of the TUSDs were verified every 4 weeks during a 20-week period with 2 further verification visits at 12-week intervals to complete the study. Data collected included the duration, intensity, and frequency of use of the TUSDs. The humidity and ambient temperature of the treatment locales were also recorded. At the initial verification, 66 percent of the TUSDs were found to be outside of tolerance limits for ultrasound intensity and 35 percent had inaccurate timers. For the duration of the study (44 weeks), 62 percent (18) of the TUSDs had to be calibrated at least twice, 38 percent (11) three times, and 31 percent (9) needed to be calibrated at least four times. Statistical analysis of the data failed to identify which factor or factors caused the deterioration in calibration of these TUSDs. Among the suggestions for improving the accuracy and safety of ultrasound treatments is that calibration verification should be made every 4 weeks and at even shorter intervals for some units. The authors suggest a follow-up study to determine the results of weekly verification.

摘要

相似文献

4
Calibration and measurement issues for therapeutic ultrasound.
Ultrasonics. 2008 Aug;48(4):234-52. doi: 10.1016/j.ultras.2007.10.010. Epub 2008 Jan 8.
5
Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines.
Ultrasound Med Biol. 2006 Sep;32(9):1423-9. doi: 10.1016/j.ultrasmedbio.2006.04.001.
6
8
Evaluation of therapeutic ultrasound equipment performance.
Ultrasonics. 2010 Jun;50(7):704-9. doi: 10.1016/j.ultras.2010.02.006. Epub 2010 Feb 14.
9

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验