易特便携式肺量计的四年校准稳定性。

Four-year calibration stability of the EasyOne portable spirometer.

机构信息

Mount Sinai School of Medicine, New York, New York, USA.

出版信息

Respir Care. 2010 Jul;55(7):873-7.

Abstract

BACKGROUND

Clinical practice guidelines recommend daily spirometer calibration checks and weekly linearity checks. The long-term stability of the volume and flow accuracy of a specific model of spirometer should be carefully characterized before modification of the frequency of calibration checks is considered for that model of spirometer.

METHODS

The EasyOne ultrasonic flow-sensing spirometer was chosen for use by the clinical centers at the 2002 inception of the World Trade Center Worker and Volunteer Medical Screening Program. The screening program quality-control procedure required that the expiratory and inspiratory volume accuracy of each spirometer be checked every day of testing, and the flow accuracy (linearity) checked every week. The calibration check results were transferred to a central database for summary.

RESULTS

Over 5,000 calibration-check results (4,109 single-speed and 1,189 three-speed) were accumulated from a total of 34 spirometers during the period February 2003 through March 2007. The mean single-speed calibration errors (and 5th-95th percentiles) were -2 mL (-80 to 70 mL) for exhalation and -10 mL (-80 to 60 mL) for inhalation. 98% of the exhalation and 97% of the inhalation calibration checks were accurate within 3.0%. There was no evidence of significant non-linearity according to the results of the 3-speed calibration checks (mean errors of -3, -5, and -6 mL at each speed).

CONCLUSIONS

The EasyOne retained inhalation and exhalation volume accuracy of better than 3% for at least 4 years. Routine multiple-speed volume calibration checks may not be necessary with the EasyOne. The acceptability and repeatability of patient efforts should be the primary focus of quality-assurance programs with spirometers that have been demonstrated to remain accurate for long periods.

摘要

背景

临床实践指南建议每天进行肺活量计校准检查和每周进行线性检查。在考虑修改特定型号肺活量计的校准检查频率之前,应仔细描述该型号肺活量计的体积和流量准确性的长期稳定性。

方法

在 2002 年世界贸易中心工作人员和志愿者医疗筛查计划启动时,选择 EasyOne 超声流量感应肺活量计供临床中心使用。筛查计划质量控制程序要求每天检查每个肺活量计的呼气和吸气量准确性,并每周检查流量准确性(线性)。校准检查结果被转移到中央数据库进行汇总。

结果

在 2003 年 2 月至 2007 年 3 月期间,从总共 34 台肺活量计中积累了超过 5000 次校准检查结果(4109 次单速和 1189 次三速)。单速校准误差的平均值(第 5 至 95 百分位数)为呼气 -2 毫升(-80 至 70 毫升)和吸气 -10 毫升(-80 至 60 毫升)。98%的呼气和 97%的吸气校准检查在 3.0%以内准确。根据三速校准检查的结果,没有证据表明存在明显的非线性(在每个速度下的平均误差为-3、-5 和-6 毫升)。

结论

EasyOne 至少在 4 年内保持了 3%以上的吸入和呼出体积准确性。对于 EasyOne,可能不需要进行常规的多速体积校准检查。对于已经证明能够长时间保持准确性的肺活量计,质量保证计划的重点应放在患者努力的可接受性和可重复性上。